Technical requirements:
1.Coating  requirements:1560.05nm  filter  coatings;operating  angle:0°~2°
2.Pass band:center wavelength:1563.05nm±2nm;FWHM≤20NM;T>92%@1563.05nm±1nm
3.Stop band:Within the stop band (400nm~1800nm wave band,excluding the pass band):Tavg<0.2%;T<0.01%@1540.56nm ±1nm  (0D4)
4.surface accuracy requirements:Testing is done with an interferometer;λ=632.8nm.
Uncoated  condition:Transmitted  wave  aberration  (including  power):superior  to  1/60  λRMS;two  surface  accuracy  (including power);:superior to 3λPV
Coated condition:Transmitted wave aberration (including power):superior to 1/40 λRMS (testing can be done with a wavefront sensor.In the case of a subaperture of φ8.8mm:RMS:superior toλ/97;λ=1563nm;4apertures at  least);parallelism:superior to 0.01mm.
| Product dimensions | 
| Nd | / | 
| Vd | / | 
| Optical uniformity | H2 | 
| Optical absorptivity | 1C | 
| Stress birefringence | 2 | 
| Fringe intensity | 1C | 
| Extent of bubble | 3 | 
| Product dimensions | 
|  | R1 | R1 | 
| Specimen precision △R | A | A | 
| Radius deviation N | / | / | 
| Partial deviation △N | / | / | 
| Eccentricity c | / | / | 
| Surface finishment MIL | Ⅲ | Ⅲ | 
| Clear aperture Do | Ф14.1 | Ф14.1 |